Discover our premium quality products sourced directly from reliable China suppliers and manufactured in our state-of-the-art factory. We specialize in delivering innovative solutions that meet the diverse needs of our clients. With a commitment to excellence and a focus on customer satisfaction, our products are rigorously tested to ensure they meet international standards. Trust in our expertise as a leading supplier from China, and elevate your business with our exceptional offerings today

    Description of veterinary Ultrasonic Scalpel

    Overview

    The ultrasonic scalpel system consists of a main unit, foot pedal, ultrasonic scalpel handle, and transducer. Ultrasound knife can generate high energy ultrasound to achieve cutting and coagulation of tissue effect.

    Application: Neutering and minor Surgery for cats, dogs and other small animals.

    Veterinary Ultrasonic scalpel is a high-frequency electrosurgical device, mainly suitable for soft tissue cutting and vascular closure and other operations. The working principle of ultrasound knife is to use ultrasound frequency generator to make the metal blade to a certain resonance frequency for high-frequency vibration, usually in the 55000Hz or so, which can make the tissue of water molecules vaporisation, protein hydrogen bond breaking, cell collapse, so as to achieve the tissue cutting or coagulation. At the same time, the veterinary ultrasonic scalpel can quickly close small blood vessels when cutting tissue, reducing bleeding and achieving almost bloodless cutting.

    ECI-UK500 Veterinary Ultrasonic Scalpel System-detail-1

    Using video

    Technical Data of Veterinary ultrasonic Scalpel

    Size Short Middle Long
    Type SRE114 SRE123 SRE136
    Length(mm) 140mm 230mm 360mm
    Frequency(KHz) 55.0
    55.0~56.0
    55.0
    55.0~56.0
    55.0
    55.0~56.0
    Power (W) Standby 2.8 2.8 2.8
    Output Gear 1MIN ~ 5MAX Gear 1MIN ~ 5MAX Gear 1MIN ~ 5MAX
    Screen size 7-inches N.W. 3.1kg Product size 240*200*140mm

    How to set the parameters of the ultrasound Knife

    The output power of the ultrasonic scalpel can be set by the “↑” and “↓” buttons on the main interface. A total of 5 gears, the maximum is 5 gears, the minimum is 1 gear, the default is the third gear, you can adjust the output power according to the actual demand.

    Click the “Settings” button to set other parameters:

    1
    Transducer test: Can test the resonance frequency and impedance of the transducer and other information.
    2
    Hand/foot control excitation: Can set the ultrasound knife excitation mode, there are three kinds.
    3
    Host work information: Can view a series of host information.
    4
    Language: Only supports English and Chinese for the time being.
    5
    Brightness: Can adjust the brightness of the screen, a total of five stops.
    6
    System Information: Can view the hardware and software version of the host computer.

    After setting the parameters, you need to click the “Back” button to return to the boot interface, and then click the “READY” button to re-initialize the cutter head.

    ECI-UK500 Veterinary Ultrasonic Scalpel System-detail-2

    Installation and removal of the cutter-head

    Align the transducer in the socket of the cutter-head on the handle, and then tighten the cutter-head can be. After the cutter-head is installed, you can connect the plug to the host computer to test whether it is installed correctly.

    Removal of the cutter-head on the veterinary ultrasonic scalpel is the reverse of the installation procedure.

    ECI-UK500 Veterinary Ultrasonic Scalpel System-detail-3
    Frequently Asked Questions
    What are the primary applications of the veterinary ultrasonic scalpel?
    It is mainly used for soft tissue cutting and vascular closure in operations such as neutering and minor surgery for cats, dogs, and other small animals.
    How does the veterinary ultrasonic scalpel achieve bloodless cutting?
    The device works at a resonance frequency of approximately 55,000Hz. This high-frequency vibration vaporizes water molecules in the tissue, breaks protein hydrogen bonds, and collapses cells, which allows it to cut tissue and quickly close small blood vessels simultaneously to minimize bleeding.
    What size options are available for the cutter-head?
    There are three options available depending on the surgery requirements: Short (140mm, Type SRE114), Middle (230mm, Type SRE123), and Long (360mm, Type SRE136).
    How do I adjust the output power of the ultrasonic scalpel?
    The output power features 5 adjustable gears (minimum 1, maximum 5). You can adjust the power using the "↑" and "↓" buttons on the main interface depending on your surgical needs. The default setting is the third gear.
    How can I test if the cutter-head is installed correctly?
    After tightening the cutter-head onto the transducer and connecting the plug to the host computer, you can run the "Transducer test" from the Settings menu. This will test and verify the resonance frequency and impedance to ensure proper installation.