What Are Small Ruminants & How to Dilute Small Ruminant Using a 100–1000 μL Adjustable Pipette
When running on-farm semen assessments, one topic our customers ask about most is small ruminant sample preparation. The biggest source of testing error comes when teams use a 100–1000 μL adjustable pipette to dilute ram semen. Poor technique skews results from your ECI-SpermPro Sperm Quality Analyzer
1.What Are Small Ruminants?
Small ruminants are sheep and goats. These animals have four-chambered stomachs and chew cud. For CASA semen analysis, ram semen holds far higher sperm concentrations than bull semen. If you load undiluted raw semen straight onto the slide, sperm stack and overlap. The ECI-SpermPro Sperm Quality Analyzer cannot reliably pick out individual sperm cells, so we use a target dilution ratio of 1:49.
Note: Deer count as ruminants but this protocol does not cover them. Alpacas are pseudo-ruminants, and this dilution method will not work for their samples.

2.How to Dilute Small Ruminant Using a 100–1000 μL Adjustable Pipette?
Two-Step Serial Dilution with a 100–1000 μL Adjustable Pipette: Steps & Why We Do It
A frequent mistake we see: technicians try to pull 20 µL of thick raw ram semen directly for single-step dilution. Ram semen is viscous, and manually pipetting such a small volume creates large sampling inaccuracies. Two-step serial dilution gives you a consistent final 1:49 dilution that works perfectly with the ECI-SpermPro Sperm Quality Analyzer.
Steps
Pre-dilution: Combine 100 µL raw semen with 400 µL specialist semen extender. Gently invert to mix. This makes a 1:5 pre-diluted blend. Each 100 µL of this mixture contains 20 µL of original ram semen.
Final dilution: Take 100 µL of your pre-mixed sample, add 900 µL fresh extender and mix gently again.
The reasoning take from the pre-diluted mix only carries 20 µL raw semen. Total extender adds up to 980 µL, giving a final ratio of 20 µL raw semen : 980 µL extender =1:49.
Besides cutting down pipetting error, serial dilution eases sudden osmotic shifts to limit sperm shock. This keeps samples stable for reading on the ECI-SpermPro Sperm Quality Analyzer.

3.Field tips
- Bring semen and extender up to 38°C before use.
- Only invert tubes slowly. Hard shaking breaks sperm tails.
- Load and run your sample on the ECI-SpermPro Sperm Quality Analyzer within 2 minutes once prepared.